Publication #2153

Reference
Drug alert no 3/2007
Name
Class 4 medicines defect information: caution in use: BOC Ltd: compressed medical nitrous oxide
Categories
Scottish Government Drug Alerts
Organization
Scottish Executive
Keywords
NHS SCOTLAND DRUGS SAFETY EFFECTIVENESS DRUG THERAPY PHARMACEUTICAL SERVICES PHARMACEUTICAL INDUSTRY MRI SCANNERS MHRA
Description
Following our drug alert no.2/2007, issued on 25 January for AZ size aluminium medical oxygen cylinders, BOC have now informed us that a small number of medical nitrous oxide AZ cylinders may also present a danger when used in MRI units. These cylinders have an old type of key operated valve fitted which contains a ferromagnetic component that makes the cylinder package potentially hazardous when used near some MRI scanners.

Contact Name
Irene Foster
Contact Address
St Andrew’s House, Regent Road, Edinburgh, EH1 3DG.
Contact Phone

Created
2007-02-05 00:00:00